Material: Stainless steel
Rated Capacities: 1t-50t
ISO 376 testing criterion
IP Rating: IP68
Suitable for force measurement device and safety controlling against overload
High precision by ISO 376, 0.5, 1, 2, high reliability
R530S has a multi-beam internal design ensuring repeatable, high accuracy sensing even in case of off-center side forces. Its design and construction of smart sensing make it an ideal load cell for a wide range of test and force measurement either as a independent sensor or embedded in production/assembly machinery. Available in a wide range of standard capacities from 1t to 50t; the R530S is designed for both tension and compression force measure-ment. Complete alloy-steel construction and hermetic sealing and high performance even in applications requiring in excess of 1 million load cycles., which ensures long-term stability and durability
Parameter | Units | Specifications | |||
Model No. | R530S | ||||
Rated capacity (Emax) | 1t/2t/3t/5t/10t/20t/30t/50t | ||||
Accuracy class1)2) | t | ±0.03 | |||
Min. dead load | kg | 0 | |||
Rated output | mV/V | 3.0 ± 0.2% | |||
Zero balance | % of Emax | ± 1 | |||
Repeatability error | % of AL3) | <±0.02 | |||
Creep; 30 minute | % of AL | <±0.02 | |||
Min. dead load output return (DR); 30 min | % of AL | <±0.02 | |||
Temp. effect on (Min. dead load output) | % of Emax/°C | <±0.002 | |||
Temp. effect on (Rated output2)) | % of AL/°C | <±0.002 | |||
Temperature range (Compensated) | °C(°F) | -10 to +40 [+14 to +104] | |||
Temperature range (Operating) | -40 to +65 [-40 to +149] | ||||
Temperature range (Safe storage) | -40 to +80 [-14 to +176] | ||||
Excitation voltage (Recommended) | V AC/DC | 5 ~15 | |||
Excitation voltage (Maximum) | 60 | ||||
Terminal resistance (Excitation) | Ω | 766± 10 | |||
Terminal resistance (Output) | 700 ± 3 | ||||
Insulation resistance @50VDC | MΩ | > 5000 | |||
Breakdown voltage | V AC | > 500 | |||
Seal type / IP rating | Hermetically welded / IP68 IP69k | ||||
Load limit (Safe) | % of Emax | 150 | |||
Load limit (Ultimate) | 200 | ||||
Material(Spring element) | Stainless steel | ||||
Material(Cable) | Ø6;4-wire;PVC | ||||
Cable length | m | 1t…30t | 50t | ||
6 | 12 | ||||
Weight; approx | kg | 1t-5t 10t-30t 50t | |||
2.5 4.0 6.0 | |||||
Fatigue life | cycles @Emax | >1,000,000 | |||
Deflection at Emax; approx | mm | < 1.0 |
1)Error due to the combined effect of non-linearity and hysteresis
2)The sum of errors due to Temperature Effect on Output comply with the requirements of OIML R60 and NIST HB44
3)AL = Applied Load